Signal to Noise ratioΒΆ

In the Huygens Software the Signal-to-Noise ratio (SNR) is treated as a regularization parameter, i.e., a means to control the sharpness of the deconvolution result.

Yet, please be aware of the effect of extreme SNR values on image deconvolution: too high SNR values can produce restoration artifacts; too low SNR values can lead to smooth structures.

Note

After a little practice it is possible to assess the SNR range for deconvolution visually. In general, the SNR should increase with the quality and amount of signal of the raw data.

There are different acquisition settings that have an impact on the quality of the recorded signal. Changing these in the acquisition system will most likely require a modification of the SNR value for deconvolution. These acquisition settings are:

  • Gain / offset
  • Time exposure / scanning velocity
  • Summing / averaging
  • Laser power
  • Spectral detection range

Because these settings are typically different for each channel HRM will require an SNR value per channel, like this:

SNRScreenshot

As a guide, these are typical SNR values for different microscopy tecniques:

  • Confocal: between 20 and 40.
  • STED: between 10 and 20.
  • Widefield: between 40 and 60.

HRM offers an automatic SNR estimator to determine suitable SNR values for the raw images. Click on SNREstimatorLink to use the SNR estimator.

In a new page you will be asked to select an SNR estimation method.

SNRAlgorithmScreenshot

The Auto method gets good estimations on images without a baseline. When working with images that contain a baseline the Legacy method can be used instead.

Next, select an image representative of the batch selected in Select images (1/5). Press Calculator22x22 for the SNR estimation:

SNREstimationScreenshot

The output of the SNR estimator shows a region of the original data next to 4 different noise simulations of the same region. In bold letters the SNR of the simulation that most resembles the original image.

By pressing Next22x22 the estimated SNR values are copied into the restoration parameter template.